Microprocessor Flame Photometer (Graphical Display) (Model No. HV-671)

Microprocessor Flame Photometer (Graphical Display) (Model No. HV-671)

HOVERLABS

  • Rs. 52,973.00
8

Salient Features :

  • Auto Gas Cutoff Facility
  • Auto Flame Failure Detection Facility
  • µP Based With Printer Interface
  • Five Element Measurement In Single Aspiration of the Sample
  • Curve Calibration Programmability using Maximum of 25 Standards
  • Data Processing Through Curve Fitting Techniques - Least Squares, Quadratic
  • Multiple Calibration Curves Saving Facility
  • Samples Data Storage Facility
  • Online Help Key
  • Printer Attachment Facility
  • Measurement in ppm & meq
  • Useful For Both Medical & Soil Testing Applications
  • Setup Storage Facility up to 1 D setups files
  • Supplied with Compressor Unit

Specifications :

  • General Range : ppm
  •                           Na: 0-100
  •                           K : 0-100
  •                           Ca: 0-100
  •                           Li : 0-100
  •                           Ba : 30-3000
  • Serum Range : meq,'I
  •                         0-200 meq/I, 1 :1 DD dil.
  •                         0-1 D meq/I, 1 :100 dil.
  •                         0-1 D meq/I, 1 :2 dil.
  •                         0-2 meq/I, 1 :5 dil. 
  •                         (all meq values are claimed with appropriate dilution)
  • Urine Range : meq,'I
  •                        0-250 meq/I, 11 DD dil.
  •                        0-100 meq/I, 1 :100 dil.
  • Sensitivity : General Mode : Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca:15 ppm Ba: 30ppm 
  • Filters : Na & K Supplied (Ca, Li & Ba are optional at extra cost) & Factory Fitted only 
  • Resolution : 0.1 ppm / 0.1 meq.
  • Reproducibility : < 1 % coefficient of variation for 20 consecutive samples using 1 D ppm Na set as Maximum Standard up to 3% CV for 20 consecutive samples of Concentrations between 50 to 199 ppm for Barium. 
  • Linearity : < 2% error when 3 ppm Na/K and 5 ppm Li are set as Maximum Std. 
  • Curve At Accuracy : ± 2% of FS 
  • Display : 128 x 64 Graphical LCD 
  • Average Time : In-built in Software 
  • Flame System : LPG & Dry Oil Free Air 
  • Detector : Photodiode 
  • Calibration : With up to 25 Standards of user choice with curve fitting software (5 for each Element). Multiple Calibration Curves can be stored in the memory 
  • Aspiration Time : 3 ml per Minute 
  • Gas Control : Adjustable with knob 
  • Flame Failure : Auto Detection 
  • Gas Cut Off : Automatic 
  • Ignition System : Auto Ignition 
  • Automiser : Axial Flow Type 
  • Dimensions : Instrument : 365 x 245 x 220 mm Approx. (LxBxH)
  • PC Interface : USB 
  • Weight : Main Unit : 8 Kg. (Approx.) 
  • Power Supply : 230 V AC ± 10%, 50 Hz 

Standard Accessories :

  • Air Compressor
  • Chimney
  • Sample Beakers
  • Manual
  • Air & Gas Tubing
  • USB Cable
  • Main Lead

We Also Recommend