Specification (PUMP)
- Flow Range : 0.001-0.999ml/min
- Delivery method : Double piston(Main & Auxiliary)
- Pump Head : 10m1 SST/Peek analytical
- Accuracy : ± 1% or ±2μl/min, whichever is greater ± 1% or ±1μl/min, whichever is greater (Micro)"
- Flow rate settings : 0-0.10 mL/min. 0.001 inc.
- 0-0.0.90 mL/min. 0.010 inc.
- 1.0 mL/min. + 0.10 inc."
- Precision : 0.25% from 0.1 mL/min. to 10 mL/min. at 20 deg C 0.25% from 0.05 mL/min. to 4 mL/min. at 20 deg C (Micro)"
- Pmax : 6000psi
- System connections : 1/16’ capilliaries
- Dimensions : 410mm x 180mm x 270mm
- Weight : 7.3kgs
- Power Input : 85 to 264 VAC, 47 to 63 Hz, 60 Watts
- Display : 320mm
- Control : Stand alone mode using front panel
Specification (UV Detector)
- Wavelength : 190-700nm
- Spectral bandwidth : 6 nm
- Wavelength Accuracy : ≤±1 nm
- Flow pool volume : 10μl
- Wavelength Reproducibility : ≤± 0.5 nm
- Optical path : 10 mm
- Dynamic Noise : ≤±2.5×10 - 5 AU (Methanol,1ml/min,254nm, 20°C)
- Static Noise : ≤± 1×10 - 5 AU (Empty pool, Response time 1 s, 20 °C)
- Dynamic Drifting : ≤±2.5×10 - 4 AU (Methanol,1ml/min,254nm, 20°C)
- Static Drifting : ≤± 1×10 - 4 AU (Empty pool, Response time 1 s, 20 °C)
- Qualitative repeatability : RSD ≤ 0.1%
- Quantitative repeatability : RSD ≤ 0.5%
- Dimensions : 160H x 260W x 350 D, (mm)
- Weight. : 6kg
- Operating Temperature : 4-40°C
- Minimum detection concentration : ≤ 1 × 1 0 - 8 g / m L ( Naphthalene/methanol solution)
- Linear range : ≥104
Salient Feature :
- Self-priming pump head requires no manual priming
- Patented “Floating Pump Seal” designed for extended seal wear
- Available with micro, analytical, semi-prep, and prep pump heads
- Programmable solvent compressibility compensation
- Easy maintenance and service for all wetted parts
- Integral prime/purge assemble for quick solvent change
|